Open Access System for Information Sharing

Login Library

 

Article
Cited 1 time in webofscience Cited 1 time in scopus
Metadata Downloads

Reduction of the base pressure of an UHV system by using turbomolecular pumps SCIE SCOPUS KCI

Title
Reduction of the base pressure of an UHV system by using turbomolecular pumps
Authors
Park, CDChung, SMSeong, DJShin, YHChung, KH
Date Issued
2007-09
Publisher
KOREAN PHYSICAL SOC
Abstract
We present the results on the reduction of the base pressure of an ultra-high-vacuum (UHV) system by using turbomolecular pumps. The UHV range is easily obtained by using a small backing pump, such as a scroll pump, a molecular drag pump, or a turbo-molecular pump. However, the extremely high vacuum range is difficult to achieve due to the low compression ratio and thermal outgassing from the inside of the turbopump. Several tests were done to overcome these difficulties. Among them, hot oxygen injection into the foreline of the pumping system is a possible way to reduce the chamber pressure by converting hydrogen to water molecule. However, a combination pumping approach turned out to be the most effective method of achieving an extremely high vacuum.
Keywords
extremely high vacuum; turbomolecular pump (TMP); combination pumping; HYDROGEN OUTGASSING RATE; STAINLESS-STEEL; HIGH-VACUUM
URI
https://oasis.postech.ac.kr/handle/2014.oak/23178
DOI
10.3938/jkps.51.912
ISSN
0374-4884
Article Type
Article
Citation
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, vol. 51, no. 3, page. 912 - 915, 2007-09