New compact and efficient local oscillator optic system for the KSTAR electron cyclotron emission imaging system SCIE SCOPUS

Title
New compact and efficient local oscillator optic system for the KSTAR electron cyclotron emission imaging system
Authors
Yoonbum NamLee, DJLee, JKim, CYun, GSLee, WPark, HK
Date Issued
2016-11
Publisher
AIP Publishing
Abstract
Electron cyclotron emission imaging (ECEI) diagnostic on Korean Superconducting Tokamak Advanced Research utilizes quasi-optical heterodyne-detection method to measure 2D (vertical and radial) Te fluctuations from two toroidally separated poloidal cross section of the plasma. A cylindrical lens local oscillator (LO) optics with optical path length (OPL) 2-2.5 m has been used in the current ECEI system to couple the LO source to the 24 vertically aligned array of ECE detectors. For efficient and compact LO optics employing the Powell lens is proposed so that the OPL of the LO source is significantly reduced from similar to 2.0 m to 0.4 m with new optics. The coupling efficiency of the LO source is expected to be improved especially at the edge channels. Results from the optical simulation together with the laboratory test of the prototype optics will be discussed in this paper. Published by AIP Publishing.
URI
https://oasis.postech.ac.kr/handle/2014.oak/37239
DOI
10.1063/1.4961290
ISSN
0034-6748
Article Type
Article
Citation
REVIEW OF SCIENTIFIC INSTRUMENTS, vol. 87, no. 11, page. 11E130, 2016-11