Spatial resolution study and power calibration of the high-k scattering system on NSTX SCIE SCOPUS

Title
Spatial resolution study and power calibration of the high-k scattering system on NSTX
Authors
Lee, WPark, HKCho, MHNamkung, WSmith, DRDomier, CWLuhmann, NC
Date Issued
2008-10
Publisher
AMER INST PHYSICS
Abstract
NSTX high-k scattering system has been extensively utilized in studying the microturbulence and coherent waves. An absolute calibration of the scattering system was performed employing a new millimeter-wave source and calibrated attenuators. One of the key parameters essential for the calibration of the multichannel scattering system is the interaction length. This interaction length is significantly different from the conventional one due to the curvature and magnetic shear effect.
URI
https://oasis.postech.ac.kr/handle/2014.oak/31226
DOI
10.1063/1.2969404
ISSN
0034-6748
Article Type
Article
Citation
REVIEW OF SCIENTIFIC INSTRUMENTS, vol. 79, no. 10, page. 10723 - 10723, 2008-10