Open Access System for Information Sharing

Login Library

 

Article
Cited 32 time in webofscience Cited 0 time in scopus
Metadata Downloads

LABORATORY CHARACTERIZATION OF AN IMAGING REFLECTOMETER SYSTEM SCIE SCOPUS

Title
LABORATORY CHARACTERIZATION OF AN IMAGING REFLECTOMETER SYSTEM
Authors
Munsat, TMazzucato, EPark, HDomier, CWLuhmann, NCDonne, AJHvan de Pol, M
Date Issued
2003-04
Publisher
IOP PUBLISHING LTD
Abstract
While microwave reflectometry has proven to be a sensitive tool for measuring electron density fluctuations in many circumstances, it has also been shown to have limited viability for core measurements and/or conditions of strong turbulence. To this end, a new instrument based on two-dimensional imaging reflectometry has been developed to measure density fluctuations over an extended plasma region in the TEXTOR tokamak. Laboratory characterization of this instrument has been performed using corrugated reflecting targets as an approximation to plasma reflections including two-dimensional turbulent fluctuations of various magnitude and poloidal wavenumber. Within this approximation, the imaging reflectometer can recover the spectral and spatial characteristics of the reflection layer lost to or otherwise inaccessible to conventional techniques.
Keywords
MICROWAVE REFLECTOMETRY; TURBULENT FLUCTUATIONS; PLASMAS; TOKAMAKS; LENGTH
URI
https://oasis.postech.ac.kr/handle/2014.oak/28775
DOI
10.1088/0741-3335/45/4/311
ISSN
0741-3335
Article Type
Article
Citation
PLASMA PHYSICS AND CONTROLLED FUSION, vol. 45, no. 4, page. 469 - 487, 2003-04