Temperature controllable HPLC column for preparative fractionation of polymers SCIE SCOPUS KCI

Title
Temperature controllable HPLC column for preparative fractionation of polymers
Authors
Im, KPark, HWKim, YChang, T
Date Issued
2008-08
Publisher
POLYMER SOC KOREA
Abstract
An HPLC column with a self-contained temperature control device was constructed for preparative temperature programmed interaction chromatography. Two Peltier plates were attached to a large bore column (120x22 mm i.d.) and the column temperature was controlled by PID mode feed back control. At a flow rate of 1.5 mL/mm, the column temperature could be increased and decreased at a rate as high as 50 degrees C/min and 10 degrees C/min, respectively, which is much faster than using a column jacket and bath/circulator. The rapid heating and cooling rates allows a high repetition rate of chromatographic fractionation. The performance of the temperature controllable column was demonstrated successfully by the fractionation of homo-polymer precursors from diblock copolymers.
Keywords
polymer fractionation; HPLC column; temperature control; Peltier device; temperature gradient interaction; chromatography; GRADIENT INTERACTION CHROMATOGRAPHY; SIZE-EXCLUSION CHROMATOGRAPHY; LIQUID-CHROMATOGRAPHY; ANIONIC-POLYMERIZATION; BLOCK-COPOLYMER; POLYSTYRENE; BEHAVIOR; METHACRYLATE)S; SEPARATION; RETENTION
URI
https://oasis.postech.ac.kr/handle/2014.oak/22514
DOI
10.1007/BF03218557
ISSN
1598-5032
Article Type
Article
Citation
MACROMOLECULAR RESEARCH, vol. 16, no. 6, page. 544 - 548, 2008-08
Files in This Item:
There are no files associated with this item.

qr_code

  • mendeley

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher

장태현CHANG, TAIHYUN
Div of Advanced Materials Science
Read more

Views & Downloads

Browse