Temperature controllable HPLC column for preparative fractionation of polymers
SCIE
SCOPUS
KCI
- Title
- Temperature controllable HPLC column for preparative fractionation of polymers
- Authors
- Im, K; Park, HW; Kim, Y; Chang, T
- Date Issued
- 2008-08
- Publisher
- POLYMER SOC KOREA
- Abstract
- An HPLC column with a self-contained temperature control device was constructed for preparative temperature programmed interaction chromatography. Two Peltier plates were attached to a large bore column (120x22 mm i.d.) and the column temperature was controlled by PID mode feed back control. At a flow rate of 1.5 mL/mm, the column temperature could be increased and decreased at a rate as high as 50 degrees C/min and 10 degrees C/min, respectively, which is much faster than using a column jacket and bath/circulator. The rapid heating and cooling rates allows a high repetition rate of chromatographic fractionation. The performance of the temperature controllable column was demonstrated successfully by the fractionation of homo-polymer precursors from diblock copolymers.
- Keywords
- polymer fractionation; HPLC column; temperature control; Peltier device; temperature gradient interaction; chromatography; GRADIENT INTERACTION CHROMATOGRAPHY; SIZE-EXCLUSION CHROMATOGRAPHY; LIQUID-CHROMATOGRAPHY; ANIONIC-POLYMERIZATION; BLOCK-COPOLYMER; POLYSTYRENE; BEHAVIOR; METHACRYLATE)S; SEPARATION; RETENTION
- URI
- https://oasis.postech.ac.kr/handle/2014.oak/22514
- DOI
- 10.1007/BF03218557
- ISSN
- 1598-5032
- Article Type
- Article
- Citation
- MACROMOLECULAR RESEARCH, vol. 16, no. 6, page. 544 - 548, 2008-08