Validation of Cuffless Blood Pressure Monitoring Using Wearable Device
SCIE
SCOPUS
- Title
- Validation of Cuffless Blood Pressure Monitoring Using Wearable Device
- Authors
- Baek, JinHyeok; Kim, JunHo; Kim, NamHo; Lee, DaeHo; Park, Sung-Min
- Date Issued
- 2018-10
- Publisher
- IEEE
- Abstract
- Ambulatory blood pressure monitoring (ABPM) has become essential to diagnose the hypertension. In the fast few decades, commercialized blood pressure monitor used oscillometric cuff-based methods which is cumbersome and unsuitable for ABPM due to the discomfort it cause to the patients. In this study, we developed the prototype of cuff-less blood pressure monitor to validate cuff-less BP estimation algorithm. Sixteen subjects were participated in the experiment to test the accuracy of the device compared to the cuff-based BP monitor using oscillometric method. Experimental results using Bland-Altman plot showed that the mean and standard deviation of error of systolic BP (SBP) and diastolic BP (DBP) were 0.3 +/- 6.6 and 1.7 +/- 9.1 respectively.
- URI
- https://oasis.postech.ac.kr/handle/2014.oak/109218
- DOI
- 10.1109/tencon.2018.8650069
- ISSN
- 2159-3442
- Article Type
- Article
- Citation
- TENCON 2018 - 2018 IEEE Region 10 Conference, page. 416 - 419, 2018-10